| It offers all the advantages of electrosurgery:
simplified haemostasis, sterile cutting, rapid
healing of the wound and a good cosmetic effect.
Its high output makes it suitable for special
procedures, e.g. in E.N.T. neurosurgery, for laproscopic
interventions, including urological and thoracic
procedures.
Microcoagulation
:
For
neurosurgical operations of for special procedures
in vascular surgery, gynaecology, etc., the application
of electrosurgical microcoagulation is of particular
advantage since the surgeon operates gently with
the bipolar technique.
Effective
vascular occlusion and microsurgery can be achieved
by this method with the very little high frequency
power. In addition, the effect of the high frequency
is limited precisely to the tissue between the
shanks of the forceps.
Salient
Features :
 |
Fully
Solid State design. |
 |
Independent
Monopolar with isolated outputs. |
 |
The
monopolar generator provides CUT, BLEND and
COAG facilities. |
 |
Operating
mode are identified by audiotones and Panel
light. |
 |
Illuminated
ON/OFF switch. |
 |
Simultaneous
CUT and COAG Command recognises COAG only
for footswitch control. |
 |
Patient
plate leads supervised during Monopolar operation.
Disconnected patient plate lead/leads will
disregard footswitch commands. |
 |
Audio-visual
indication for patient plate lead faults. |
 |
Natural
convection cooling with silent operation. |
 |
Bipolar
generator has auto handswitching with delay
or footswitch control. |
Standard
Accessories :
 |
Surgiccal
handle with cable |
 |
Electrodes
set |
 |
Patient
Plate with cable |
 |
Double
Pedal Footswitch with cable |
 |
Mains
cable |
 |
Rexin
Cover |
Optional
Accessories :
 |
Monopolar
forceps with cord and jack |
 |
S.S.
Paediatric Patient plate with cable |
 |
Mobile
trolley for above |
 |
Servo
controlled voltage stabilizer |
 |
Pencil
handswitch with cable and jack |
Strongly
recommended areas with AC line Voltage fluctuations.
High
Frequency Electrosurgical Unit with Bipolar facility.
It gives characteristic of electro-surgical current
for effective cutting, co-agulation used for all
major and minor operations. Has facility of monopolar
and bipolar current.
Technical
Data :
Output
Monopolar Generator |
: (Frequency
500 ± 25 Khz).
: CUT Control Full Scale (PURE) 375 ±
10 W, 500 Q load.
: CUT Control Full Scale (BLEND) 250 ±
10 W, 500 Q load
: (continuous blending through Hemostasis control).
|
| Bipolar
Generator |
: COAG
Control 125 ± 15 W (Frequency 650 ±
25 Khz) |
| Output
Level |
: 70
± 5W, 125 £2 load |
| Power |
: 230
V ± 5%, 50 Hz |
| Size
|
: 480
x 290 x 180mm |
| Weight
|
: 15
Kg |
Applications
:
For
all types of MAJORSURGERY includes under water
cutting T.U.R. and for bipolar micro CO-AGULATION.
|